November 22, 2026 — React Health has issued a voluntary medical-device correction affecting 7,533 Luna G3 BPAP systems distributed in the United States and Puerto Rico because certain devices may record or transmit inaccurate therapy-pressure statistics when the Targeted VT feature is enabled. The Food and Drug Administration posted the company’s updated notice November 19.

The affected device is the Luna G3 BPAP System, model LG3800-30VT. According to React Health’s November 17 notice, the problem involves the data stored on the device’s SD card and transmitted to its cloud-based management platform. With Targeted VT enabled, recorded pressure statistics may not reflect the pressures actually delivered, potentially misleading clinicians who review historical reports and adjust treatment settings. The company says devices without Targeted VT enabled do not experience this particular issue.

What patients should do

Do not independently change prescribed settings or discontinue respiratory therapy. React Health advises patients using the affected model with Targeted VT to contact their treating clinician or durable medical equipment provider promptly to determine whether their device is affected and arrange any necessary therapy adjustments or corrective action. Patients should continue prescribed therapy until their provider gives further instructions.

React Health reported 15 complaints, including two serious adverse reports, associated with this issue as of its notice. The company developed firmware version G3-2.00.80 to correct the reporting problem and says providers can coordinate installation. The firmware does not retroactively correct inaccurate historical pressure data, which clinicians should interpret cautiously.

Instructions for clinicians and equipment providers

The company’s guidance instructs clinicians and equipment providers to discontinue the Targeted VT feature on affected devices while coordinating appropriate alternative patient support; review affected inventory; and arrange the firmware correction. These instructions are directed to healthcare professionals and should not be interpreted as a recommendation for patients to stop therapy without medical supervision.

Patients and providers can consult the FDA recall posting and React Health’s detailed correction notice, including affected serial numbers and provider instructions. React Health lists 863-226-6285 for provider support. The FDA republishes the company’s announcement as a public service; its posting does not constitute an independent FDA endorsement of the company’s statements.

Sources: FDA, posted November 19, 2026; React Health, updated November 17, 2026.